SonoSite Edge II Troubleshooting Guide

Here are a couple of initial troubleshooting steps for the Sonosite Edge II you can take to ensure your study crosses over:

1) Ensure that the current study has been appropriately ended. This can be done by navigating to the patient information screen by clicking on Patient and then selecting New/end. This will start the uploading process for your study.

2) If your study has been appropriately ended, next check to see if it has been "archived" (sent to Synchronicity). You can do this by navigating to the past study list. Get to this by clicking on Review (blue arrow) on the main control panel. This will bring up the list of all studies stored on the machine.

3) From the patient list, look to see if your study has a check mark next to it in the rightmost upload column (green box). If so, it has been sent to Synchronicity. If there is no check mark, you will need to re-submit the study as it has not been uploaded.

4) If there is no check mark next to the study, select the study by clicking on it using the mouse/trackpad. Once you have selected all the studies you wish to archive, then press "Archive" (green arrow) to re-submit to Synchronicity.

5) Remember to leave the machine open and powered on after performing the above steps to allow the files to upload. Wait a few minutes for your study to upload to Synchronicity and then log in and proceed with workflow as normal.

If you are unable to resolve the issue on your own, please call the IUH Service Desk and open a ticket.
Service Desk Phone Number: 317-962-2828